Maintenance
Use the following procedure to clean the printhead.
1.
Dip cotton swabs in ethyl alcohol and wring out the excess solution,
NOTE:
Do not use products that can harm the heating element, such as sandpaper.
2.
Open the printer door.
3.
Gently wipe the heating element with the cotton swabs.
NOTE:
The printhead gets hot when recording. Do not touch the thermal printhead
directly.
Avoid unnecessary force when handling the printhead.
4.
Reinsert the paper and close the printer door when the heating element is
completely dry.
NOTE:
Use only original GE Healthcare writer paper. This paper has a special coating that
prevents electrostatic buildup and contamination and debris collection on the
printhead. Using other paper may result in recordings of poor quality. Use of other
papers may wear out the printhead prematurely and may void the warranty.
Removing the Printhead
Use the following procedure to remove the printhead.
1.
Before beginning any steps, follow the instructions in
FRU Replacement” on page 73
2.
Remove the printer module assembly.
See
“Removing the Printer Module Assembly” on page 94
for instructions.