Service Setup
Status
Description
Battery Life Remaining
The estimated time remaining based on the
present current draw.
Because no printing is occurring during this
status check, this time is calculated based on
using the system for display purposes only.
Actual remaining battery life is less when
using the system for both display and printing
purposes.
Battery Charge Remaining
According to the battery specifications you
should see 3:00 (three hours) or less of
Battery
Life Remaining
(display time) with 100%
Battery
Charge Remaining
.
If the numbers displayed in this window are
less, you may require more frequent battery
charging or you may wish to replace the battery.
See
“Replacing the Battery” on page 74
for
information on how to replace the battery.
NOTE:
• This system is internally powered with lithium ion batteries with different
levels of safety protection. It is not recommended to charge Lithium ion
batteries above 45°C (113°F). If the device is operating at a high ambient
temperature, and the battery temperature exceeds the recommended
operating conditions, charging stops automatically. Under such
conditions, turn off the device and let it cool before restarting charging.
• You can extend the battery life by reducing the brightness on the display.
• To decrease the brightness, press
Shift
+
Cursor Down
.
• To increase the brightness, press
Shift
+
Cursor Up
.
4.
Note the battery status information and press
Cancel
to close the
BATTERY
STATUS
window.
Testing the Writer
Use the
Writer Test
to determine if the writer is working properly.
NOTE:
Before performing the
Writer Test
, be sure that thermal paper is loaded properly.
Refer to your system’s Operator Manual for instructions on loading paper.
1.
Open the
Diagnostic Tests
window as described in
page 46
2.
Select
Writer Test
.
The following window opens.
52
MAC™ 2000 ECG Analysis System
2053535-003C
Summary of Contents for MAC 2000
Page 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Page 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Page 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Page 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Page 183: ......