Maintenance
3.
Holding the printer module assembly vertically, remove the fastener securing
the printer motor cover to the printer module assembly, as shown in the
following photograph, and lift off the printer motor cover.
4.
Remove the two fasteners securing the printer motor to the printer module
assembly, as shown in the following photograph, and lift off the printer motor.
5.
Remove the three (3) C-clips securing the gears to the printer module assembly
and remove the gears, as shown in the following photographs.
Replacing the Printer Motor and Gear Assembly
Use the following procedure to replace the printer motor and gears.
1.
Attach the gears to the printer module assembly and secure with the three (3)
C-clips, as shown in the following photographs.
96
MAC™ 2000 ECG Analysis System
2053535-003C