Service Setup
3.
Select
Start Test
.
The following window opens.
4.
Press the
right arrow
on the
Trimpad
repeatedly to move the color bars
horizontally across the screen.
5.
Verify that the color band pattern (red, green, blue, white) scrolls across the
screen.
Pass the test if the pattern is replicated without discoloration.
6.
Press the first function key on the left to switch to horizontal color bars.
7.
Press the
down arrow
on the
Trimpad
repeatedly.
8.
Verify that the color band pattern (red, green, blue, white) scrolls down the
screen.
Pass the test if the pattern is replicated without discoloration.
9.
Press the first function key again to cycle through the solid color pane (red,
green, blue, white).
10.
For each pane, check for black pixels.
Pass the test if no more than 4 black pixels are observed on any single color
pane.
NOTE:
A black pixel observed on one pane will probably be observed on every pane.
11.
Press
Esc
or
Enter
when the test is done.
The following window opens.