Maintenance
Replacing the Printer Door Sensor
Use the following procedure to replace the printer door sensor on the printer module
assembly.
1.
Snap the printer door sensor into the printer door sensor cover, as shown in the
following photograph.
2.
With a pair of needle nose pliers, gently snap the printer door sensor cover into
place on the printer module assembly, as shown in the following photograph.
3.
Position the printer motor cover over the printer motor and secure with its
fastener, as shown in the following photograph.
4.
Reattach the printer module assembly to the bottom cover.
See
“Replacing the Printer Module Assembly” on page 95
for instructions.
5.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
Display Assembly
The following procedures describe how to remove and replace the display assembly.
108
MAC™ 2000 ECG Analysis System
2053535-003C
Summary of Contents for MAC 2000