Maintenance
Item
Description
1
Top Cover
2
Chassis
3
Bottom Cover
4
Power Supply
5
Main PWA
6
Display
7
Keypad
8
Patient Cable
9
Internal Modem
10
Battery
11
Printer Assembly
12
KISS Pump
Preparing the System for FRU Replacement
Prior to performing any disassembly procedures, perform the following steps:
NOTE:
Take strict precautions against electrostatic discharge damage while replacing field
replaceable units.
1.
Store and print your device settings so that you can restore them or manually
reconfigure the system when you are done.
For information on storing your settings, see
For information on printing your settings, see
“Printing the System Setup Report”
on page 121
.
2.
Power down the system.
3.
Disconnect the device from the AC wall outlet (1).
4.
Disconnect the power cord from the rear of the device (2).
5.
Disconnect the patient cable from the device.
6.
Disconnect all cables from the back panel.
2053535-003C
MAC™ 2000 ECG Analysis System
73