Maintenance
Replacing the Power Supply Assembly
Use the following procedure to replace the power supply.
1.
Position the new power supply assembly onto the power supply base.
2.
Insert the four (4) screws to secure the power supply assembly to its base, as
shown in the following photograph.
3.
Attach the power supply harness to the PWA, as shown in the following
photograph.
4.
Secure the ground to the PWA using the fastener, as shown in the following
photograph.
5.
Replace the top cover assembly.
See
“Replacing the Top Cover” on page 79
for instructions.
6.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
7.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
Power Inlet Module
The following procedures describe how to remove and replace the power inlet module.
92
MAC™ 2000 ECG Analysis System
2053535-003C