Maintenance
d.
Select
Device Settings
.
The following window opens.
e.
In the
Serial Number
field, type the serial number of the device.
Refer to
“Serial Number Format” on page 29
for information about
the serial number.)
f.
Select the appropriate language from the
Keyboard
list.
g.
Press
Save
.
11.
Reactivate the purchased options.
Refer to the operator’s manual for your system for instructions on activating
the options.
12.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
KISS Pump Assembly (Option)
The following procedures describe how to remove and replace the KISS pump
assembly and tubing.
Removing the KISS Pump Assembly
Use the following procedure to replace the KISS pump assembly.
1.
Remove the top cover.
See
“Removing the Top Cover Assembly” on page 75
for more information.
2.
Detach the KISS pump tubing from the locations identified in the following
photograph.
126
MAC™ 2000 ECG Analysis System
2053535-003C