K415 Issue No. 1
46
Internal Pressure Range
Use the following procedure for calibrating the internal pressure range.
Note:
If calibrating the hydraulic calibrator version, the calibrator must first be primed as
detailed on pages 59 to 66.
(1)
Connect the outlet port of the instrument to a pressure standard.
(2)
Allow the instrument’s temperature to stabilize for a minimum of 1 hour.
(3)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu and then
PRESSURE INT
.
(4)
Apply zero pressure and store the zero point:
(5)
Close the vent valve, apply full-scale pressure and store the full-scale point:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit
CALIBRATION
and
SETUP
modes.
(7)
Check calibration by applying test pressures in Table 1, page 44.
VOLTAGE IN
Calibration Terminee
App ENTER pour continuer
Memoriser point PE ?
OUI
NON
m = 2.784670e-04
c
= -1.02
Applique
Mesure
0.0000
524293
20.000
704637
CALIBRE PRESSION INT
GAMME
20.000
bar g
CALIBRE PRESSION INT
GAMME
20.000
bar g
RESULTATS
Calibration
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...