27
K415 Issue No. 1
mA Ramp
Press the
OUTPUT
key and select
mA RAMP
as shown previously in mA Step.
• Define ramp required by entering
START
and
END
current values as shown below:
• For current sourcing applications, turn on the 24V supply as shown on page 28. For
current sinking applications, connect an external supply as shown on page 25 and
leave the 24V supply switched
OFF
.
• Press
START
(F1) to run the program. A status display
CHECK LOOP
indicates a fault
in the external loop i.e. supply fault or open circuit.
Note:
The ramp cycle (min to max or max to min), is approximately 60 seconds.
• Press
STOP
(F1) when running to stop at any point. Press
CONTINUE
(F1) to resume
from point of pause or
RESET
(F2) to return to start point.
ENTRER VALEUR
ENTRER VALEUR
bar
DEBUT
DEBUT
SORTIE
FIN
mA
ARRET
FCT. : CONFIG.
MODE ENREG. RAPIDE
bar
DEBUT
DEBUT
SORTIE
FIN
mA
ARRET
FCT. : CONFIG.
MODE ENREG. RAPIDE
24V OFF
24V OFF
ENTRER VALEUR
ENTRER VALEUR
bar
DEBUT
DEBUT
SORTIE
FIN
mA
ARRET
FCT. : CONFIG.
MODE ENREG. RAPIDE
bar
ARRET
DEBUT
SORTIE
FIN
mA
FCT. : CONFIG.
MODE ENREG. RAPIDE
EN COURS
24V OFF
24V OFF
ENTRER VALEUR
ENTRER VALEUR
bar
ARRET
DEBUT
SORTIE
FIN
mA
ARRET
FCT. : CONFIG.
MODE ENREG. RAPIDE
bar
CONTINUE
RESET
DEBUT
SORTIE
FIN
mA
FCT. : CONFIG.
MODE ENREG. RAPIDE
PAUSED
24V OFF
24V OFF
Advanced Task
Select Output
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...