The calibration procedures require the following test equipment.
*ppm = parts per million
Table 1 Internal/External Pressure Transducer Verification
PC* = Pressure Calibrator
HC* = Hydraulic Calibrator
LP
* = Low Pressure Calibrator
Test Equipment and Instrument Parameter/Range
Calibration
Equipment
Accuracy
Calibration
Uncertainty
Digital Voltmeter - 5V input
±30 *ppm ±1 digit
±10 *ppm ±5µV
Digital Voltmeter - 50V input
±45 *ppm ±1 digit
±11 *ppm ± 110 µV
Digital milliammeter - 55mA input
±150 *ppm ±4 digit
±100 *ppm ± 1nA
Digital milliammeter - 24mA output
±150 *ppm ±4 digit
±160 *ppm ± 1nA
Precision thermometer - ambient temperature
±0.2°C
± 0.1°C ±1 digit
Deadweight tester - pressure internal/external
Class A deadweight
<0.025%
Nominal applied value as a% of
±full-scale
Permissible deviation
PC*, HC* and Indicator
Versions
LP
*
Versions and
external LPE 9400
sensors
0% (10% for absolute ranges)
± 0.025% FS
± 0.05% span
20%
± 0.025% FS
± 0.05% span
40%
± 0.025% FS
± 0.05% span
60%
± 0.025% FS
± 0.05% span
80%
± 0.025% FS
± 0.05% span
100%
± 0.025% FS
± 0.05% span
80%
± 0.025% FS
± 0.05% span
60%
± 0.025% FS
± 0.05% span
40%
± 0.025% FS
± 0.05% span
20%
± 0.025% FS
± 0.05% span
0% (10% for absolute ranges)
± 0.025% FS
± 0.05% span
Calibration
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...