K415 Issue No. 1
50
(5)
Set the variable supply voltage to 50V ± 0.1V and enter the measured full-scale
voltage:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument calibration by applying the test voltages given in Table 2 to
the voltmeter (after both voltage ranges have been calibrated).
(8)
Disconnect the calibration/test equipment.
Table 2 Electrical Voltage Input Calibration Tolerances
Voltage Calibration 50V Range Verification Tolerances Voltage Calibration 5V Range Verification Tolerances
Applied Voltage
Permissible Deviation
Applied Voltage
Permissible Deviation
-50
± 0.05% rdg, ± 0.004% FS
-5
± 0.05% rdg, ± 0.004% FS
-40
± 0.05% rdg, ± 0.004% FS
-4
± 0.05% rdg, ± 0.004% FS
-30
± 0.05% rdg, ± 0.004% FS
-3
± 0.05% rdg, ± 0.004% FS
-20
± 0.05% rdg, ± 0.004% FS
-2
± 0.05% rdg, ± 0.004% FS
-10
± 0.05% rdg, ± 0.004% FS
-1
± 0.05% rdg, ± 0.004% FS
0
± 0.05% rdg, ± 0.004% FS
0
± 0.05% rdg, ± 0.004% FS
10
± 0.05% rdg, ± 0.004% FS
1
± 0.05% rdg, ± 0.004% FS
20
± 0.05% rdg, ± 0.004% FS
2
± 0.05% rdg, ± 0.004% FS
30
± 0.05% rdg, ± 0.004% FS
3
± 0.05% rdg, ± 0.004% FS
40
± 0.05% rdg, ± 0.004% FS
4
± 0.05% rdg, ± 0.004% FS
50
± 0.05% rdg, ± 0.004% FS
5
± 0.05% rdg, ± 0.004% FS
CALIBRE TENSION
GAMME +/- 50V
Memoriser point PE ?
OUI
NON
CALIBRE TENSION
GAMME +/- 50V
RESULTATS
m = 1.395879e-05
c = -7.00
Applique
Mesure
0.0000
526380
49.984
873988
Calibration Terminee
App ENTER pour continuer
CALIBRE TENSION
GAMME +/- 50V
Appliquer PE nominale
App sur ENTER quand
la valeur est stable
REGLER VALEUR
CHANGER
VALEUR
DEMARRER
CALIBRE TENSION
GAMME +/- 50V
Entrer val appliquee:
CHANGER
VALEUR
DEMARRER
CALIBRE TENSION
GAMME +/- 50V
Enter val appliquee:
Calibration
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...