K415 Issue No. 1
52
(5)
Set the input current to 55 ± 0.1mA and enter the measured full-scale input
current:
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument calibration by applying the test currents given in Table 3 to
the milliammeter.
(8)
Disconnect the calibration/test equipment.
Table 3 Electrical Current Input Calibration Tolerances
Current Calibration, 55mA Range Verification Tolerances
Applied Current (mA)
Permissible Deviation
-55
± 0.05% rdg ± 0.004% FS
-45
± 0.05% rdg ± 0.004% FS
-35
± 0.05% rdg ± 0.004% FS
-25
± 0.05% rdg ± 0.004% FS
-15
± 0.05% rdg ± 0.004% FS
-5
± 0.05% rdg ± 0.004% FS
0
± 0.05% rdg ± 0.004% FS
5
± 0.05% rdg ± 0.004% FS
15
± 0.05% rdg ± 0.004% FS
25
± 0.05% rdg ± 0.004% FS
35
± 0.05% rdg ± 0.004% FS
45
± 0.05% rdg ± 0.004% FS
55
± 0.05% rdg ± 0.004% FS
Calibration
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...