K415 Issue No. 1
28
mA Value
Press the
OUTPUT
key and select
mA
VALUE
from the output menu. The procedure is
shown below:
• For current sourcing applications, turn on the 24V supply as shown below. For
current sinking applications, connect an external supply as shown on page 25 and
leave the 24V supply switched OFF.
• Use
up
and
down
cursor keys to adjust output current level. While the loop is
made, a status display indicates
OK.
A status display
CHECK LOOP
indicates a fault
in the external loop i.e. supply fault or open circuit.
24 Volt
Press the
OUTPUT
key and select
24 VOLT
from the Output menu. The procedure is
shown below:
Note:
The 24V selection is located on the second page of output functions.
Note:
To conserve battery power, keep the 24V supply switched off when not in use (even
if the output is unloaded).
CHOISIR SORTIE
REGLER NIVEAU
SELECT SORTIE
PRESSION
INT
PRESSION EXT
PAS mA
RAMPE mA
VALEUR mA
bar
SORTIE
NOUV VAL
mA
OK
FCT. : CONFIG.
MODE ENREG. RAPIDE
24V OFF
OUTPUT
VALIDER
SORTIE
SELECTIONNER
SORTIE
SELECT SORTIE
24 VOLT
OFF
PAS DE SORTIE
SELECT SORTIE
24 VOLT
ON
PAS DE SORTIE
Advanced Task
Select Output
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...