K415 Issue No. 1
Voltage Input Range (5 Volts)
Use the following procedure for calibrating the internal 5 Volt range.
(1)
Switch the instrument on, enter
CALIBRATION
mode and select
INT RANGES
from
the
CALIBRATION
menu as shown on page 45.
(2)
Select 5V range for calibration:
(3)
Link the voltage input terminals with a short lead and enter the zero point:
(4)
Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
SELECTIONNER
OPTION CAL
SELECTIONNER
OPTION CAL
SELECT. GAMME
50 VOLTS
5 VOLTS
CALIBRATION
TENSION
ENTR. COURANT
SORT. COURANT
PRESSION INT
TEMPERATURE
Calibration