K415 Issue No. 1
54
(4)
Measure the 10% full-scale output current and enter the value measured on the
external milliammeter:
(5)
Measure the 90% full-scale output current and enter the value measured on the
external milliammeter
(6)
Press the
ENTER
key to accept the calibration. Press the
EXIT
key four times to
quit the
CALIBRATION
and
SETUP
modes.
(7)
Verify the instrument’s output current calibration by setting up the loop test
currents as shown in Table 4 and checking the set values against the
milliammeter standard.
(8)
Press the
OUTPUT
key, select
24 VOLT
and turn the 24V supply to
ON
. Proceed as
follows:
Measure 90% FS - output
Press ENTER key when
applied value is stable
Store Cal point?
YES
NO
Enter Measured Current:
RUN
:
Calibration