Current to Pressure Converter (I-P) Task
A - External pressure supply
B - Pressure to current device (24V)
• Use the
up
and
down
cursor keys to adjust the loop current to the required
value. Alternatively, press
ENTER
and use cursor keys to enter a finite value. Cursor
keys can then be used to nudge the output either up or down. If required, change
pressure units with INPUT key.
Pressure Switch Test (P-SWITCH) Task
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Pressure switch under test
• Contact state will be shown on display. When contacts close, buzzer sounds.
• To run switch test, close vent valve and press the
RUN
(F1) key.
• Using the hand-pump, increase the applied pressure to just below the switch
operating point. Screw the volume adjuster in until the switch operates (the display
then shows the operating pressure of the switch).
• Reduce pressure until the switch releases (indicated by the switch symbol). The
display then shows the release pressure and the hysteresis value.
P
I
+ (24V)
-
bar
F2
F1
SORTIE
NOUV VAL
mA
OK
FCT. : I-P
24V OFF
MODE ENREG. RAPIDE
DPI 615
B
A
bar
F2
F1
DEMARRER
ETAT CONTACT
A
B
FCT. : P-PRESSOS
24V OFF
MODE ENREG. RAPIDE
DPI 615
C
Taking Measurements
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...