K415 Issue No. 1
24
Flow Function
To apply the flow function to a selected input, enable
FLOW
from the process
menu and press
ENTER
. The square root symbol is displayed beside the input
value to indicate that the
FLOW
function is active:
To cancel FLOW, press INPUT and turn function
OFF at the process menu.
%Span
To convert a selected input display from a numerical value to a percentage of
full-scale reading, enable
SPAN
from the process menu and press F1 (SETTINGS)
to provide access to the span definition parameters. Two span definitions are
required,
Zero
and
Full Scale
.
To leave span at current setting, press
EXIT
.
To define zero and full-scale settings, proceed as follows:
To cancel
%SPAN
, press
INPUT
and turn function OFF at the process menu.
bar
V
FCT. : CONFIG.
24V OFF
MODE ENREG. RAPIDE
SELECTIONNER
VALEUR
SELECTIONNER
VALEUR
DEFINE ECH.
VALEUR AFFICHEE
V
50.000
V
0% =
100% =
ENTER UNE VALEUR OU F2
POUR SELECTIONNERVALEUR
ACTUELLE OU EXIT POUR
QUITTER SANS MODIFICATIONS
DEFINE ECH.
VALEUR AFFICHEE
V
V
0.000
0% =
100% =
_
_
ENTER UNE VALEUR OU F2
POUR SELECTIONNERVALEUR
ACTUELLE OU EXIT POUR
QUITTER SANS MODIFICATIONS
CHANGER
VALEUR
CHANGER
VALEUR
LECTURE
ACTUELLE
LECTURE
ACTUELLE
Advanced Task
Process Functions
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...