Pressure Transmitter (P-I) Task
Select the P-I task from the task menu and connect the Unit Under Test (UUT) to the
calibrator as shown below.
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Pressure to current 24V device
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Voltage Output Pressure Transmitter (P-V) Task
Select the P-V task from the task menu and connect the UUT to the calibrator as shown
below. Voltage output transducers with a 10 V supply and outputs of ±10 V can be
calibrated using the 10V output sockets (supply) and the transducer output connected to
a voltmeter (refer to Method on page 13 for a test procedure).
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Excitation 10V
• If required, select the output units as described on page 10.
• If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
P
I
+ (24V)
-
bar
mA
F2
F1
A
B
FCT. : P-I
24V OFF
MODE ENREG. RAPIDE
DPI 615
C
F2
F1
P
V
-
+
+
-
+
-
bar
V
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
A
B
C
Taking Measurements
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...