K415 Issue No. 1
Typical Calibration Set-up (Pressure to Voltage)
Connect a device under test to the instrument as shown below:
A - External pressure source (indicator instruments only)
B - Pressure regulator
C - Excitation 10V
General procedure
• Use the hand-pump to pressurize the system to the required level as indicated on the
display. Allow the display to settle and screw the volume adjuster in or out as a fine
adjustment to the required pressure. Record the input (e.g.)
Voltage
, reading at each
applied pressure.
Zero Display Reading
Both the input and output readings can be set to zero using the
ZERO
key and if the
displayed reading is within 5% of zero. To zero either the
INPUT
or
OUTPUT
displays,
proceed as follows:
F2
F1
P
V
-
+
+
-
+
-
bar
V
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
A
B
C
ZERO
bar
V
OUI
TENSION A ZERO ?
NON
bar
V
FCT.: STANDARD
FCT.: STANDARD
ZERO
bar
V
OUI
CAPTEUR INTERNE A ZERO ?
NON
bar
V
FCT.: STANDARD
FCT.: STANDARD
ZERO
ENTRÉE
ZERO
ENTRÉE
ZERO
SORTIE
ZERO
SORTIE
Getting Started
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...