75
K415 Issue No. 1
Calibration
The calibration routines for the DPI 610 and DPI 615 Low Pressure calibrators are as
described in calibration page 43, with the exception of the internal pressure range. The
calibration procedure for the internal pressure range requires the application of five test
pressures as detailed below.
Note:
Refer to page 43 for the general procedures of calibrating the instrument. Refer to
Table 1, page 44 for the transducer verification.
Calibrate Internal Pressure Range
To calibrate the internal pressure range:
Connect the outlet port of the instrument to a pressure standard.
Switch the instrument
ON
.
Allow at least one hour for the temperature of the instrument to stabilize.
Select
SETUP
and enter the
SETTINGS
menu.
Note:
SETUP cannot be selected from BASIC mode.
Select
CALIBRATION
from the
SETTINGS
menu and enter the calibration
PIN
number (refer to page 45 for PIN number details).
Select
PRESSURE INT
from the
CALIBRATION
menu (refer to page 46).
Apply negative full-scale pressure to the instrument. Store the negative full-scale
point:
Apply negative, half full-scale pressure to the instrument and store the negative
half full-scale point:
Entrer val appliquee:
Appliquer valeur -ve PE
App sur ENTER quand la
la valeur est stable
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
NON
OUI
CHANGER
VALEUR
Valider point -ve P.E.?
REGLER VALEUR
Entrer val appliquee:
Appliquer 50% valeur -ve
App sur ENTER quand la
la valeur est stable
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
CALIBRE PRESSION INT
GAMME 2.5000 mbar L
NON
OUI
CHANGER
VALEUR
Valider 50% point -ve?
REGLER VALEUR
Low Pressure Calibrator Versions
Summary of Contents for Druck DPI 610
Page 1: ...Druck DPI 610 615 Portable Pressure Calibrator Series User manual K415 GE Sensing ...
Page 2: ......
Page 10: ...K415 Issue No 1 viii ...
Page 69: ...59 K415 Issue No 1 Hydraulic Calibrator Versions ...
Page 70: ...K415 Issue No 1 60 This page intentionally left blank ...
Page 76: ...K415 Issue No 1 66 This page intentionally left blank ...
Page 77: ...67 K415 Issue No 1 Low Pressure Calibrator Versions ...
Page 78: ...K415 Issue No 1 68 This page intentionally left blank ...
Page 87: ...77 K415 Issue No 1 Appendix 1 DATA LOG FILE EXAMPLE ...
Page 88: ...K415 Issue No 1 78 This page intentionally blank ...