B30 Patient Monitor
10-2
Messages
Table 10-1
Messages
Message
Explanation
What to do
Air leakage
Air leak in cuff or hose.
•
NIBP: Check all connections and
test tightness using Venous Stasis.
Alarms acknowledged (from
Central)
Silenced alarms remain silent. New
alarms will have an audible sound.
(Can be done using the Central).
•
If required, turn on the alarms
through
Alarms Setup
– Audio
ON/OFF – Activate Alarms
.
Alarm setup changed from
Central
Alarm limits or arrhythmia alarm
priorities have been changed using
the Central.
•
Check the alarm limits and the
arrhythmia alarm priorities; see
and
.
Alarms silenced from Central Alarms have been silenced using the
Central.
•
If required, turn on the alarms
through
Alarms Setup
– Audio
ON/OFF – Activate Alarms
.
Apnea
No breath detected for 20 seconds.
•
Check the patient status.
•
Check the ventilator and breathing
circuit.
Apnea deactivated
The case has just been started or the
measurement has just been activated,
and apnea alarm is not active yet. It
will activate after 5 breaths.
•
The message will disappear after
the monitor detects three breaths.
Artifacts
Some artifacts (muscle activity, eye
movements etc.). may disturb certain
measurements.
•
Calm the patient since patient
movements, shivering, deep
breathing, arrhythmia or irregular
beats may cause some
measurements to fail.
•
If applicable, start a new
measurement.
Asystole
No QRS detected in ECG.
•
Check the patient status.
•
Check the electrodes.
Battery low
About 20 minutes of battery operation
time left.
•
Replace the battery or connect the
monitor to power outlet.
Brady
HR is equal to or below the lower
alarm limit.
•
Check the patient status.
Calibr. error
Unsuccessful gas calibration.
•
Perform a new calibration.
Check D-fend
The water trap is not attached.
•
Check that the water trap is
properly attached to the module. If
the problem persists, contact
authorized service personnel.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......