B30 Patient Monitor
5-4
Changing alarm options
1.
Press the
Monitor Setup
key, select
Install/Service
and enter the password.
2.
Select
Installation - Alarm Options
.
•
Select
Show limits
and
YES
to show alarm limits in digit fields.
NO
is the default.
•
Select
Show Audio ON/OFF
(default) and
YES
to enable alarm silencing. Selecting
NO
disables silencing options in the
Audio ON/OFF
menu in
Alarms Setup
.
NOTE: The
Show Audio ON/OFF
setting should be changed only by the system
administrator.
•
Select
Alarm Light
and
YES
(default) to enable the alarm light or
NO
to disable it.
•
Select
Latching Alarms
and
YES
to keep alarm messages on screen until
Silence
Alarms
key is pressed.
NO
is the default.
•
Select
Reminder Volume
and adjust the volume of the audible alarm reminder tone.
•
Select
Alarm Tones
and
ISO
,
ISO2
or
General
.
Other alarm settings (alarm limits, alarm light brightness and alarm volume) can be changed in
the
Alarms Setup
menu. To make the changes permanent, save them in user modes, see
”Saving changes in user modes.”
Changing printer settings
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Installation - Printer
.
•
Select
ECG Printout Type
and
2x6-25 mm/s
(2 columns, 6 lines, 25 mm/s; default),
2x6-50 mm/s
(2 columns, 6 lines, 50 mm/s) or
3x4-25
mm/s
(3 columns, 4 lines, 25
mm/s).
•
Select
Printer Connection
and
Net1…Net16
/
None
(default).
•
Select
Paper Size
and
A4
(default) or
Letter
.
NOTE: Network printer only.
Other printer settings can be changed in the
Print & Record
menu. To make the changes
permanent, save them in user modes.
Changing the monitor settings
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Installation - Monitor Settings
.
•
Select
Parameter Settings
and select:
Humidity compensation type:
CO2 Numbers - Dry
(default),
Wet
.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......