B30 Patient Monitor
14-2
Display of invasive blood pressure
Figure 14-1 InvBP display
(1) Invasive blood pressure waveforms with zero and reference lines
(2) Selected
pressure
label
(3) Selected
pressure
scale
(4) Field for messages and alarm limit settings
(5) Systolic, diastolic and mean pressure values of invasive blood pressures
You can have a combined display of all those waveforms that are selected on the screen. This
combined display uses the whole waveform field area and the same zero line for all
waveforms.
To select:
1.
Press the
Monitor
Setup
key and select
Screen Setup
.
2.
Select
Waveform Fields
.
3.
To combine all waveforms in one field, select
Combine Pressures
and
Yes
.
Patient connections
1.
Connect the pressure transducer to the transducer adapter cable. Connect the cable to
the red connector in the module, or to the dual invasive blood pressure adapter cable.
NOTE: Invasive pressures need to be zeroed after reconnecting the pressure
transducer or cable, and whenever the patient’s position is changed. The transducer is
always leveled to the mid right atrium.
2.
Prepare the transducer kit according to the manufacturer’s instructions. Mount the kit
with the transducer zeroing port at mid-heart level.
3.
Ensure that there is no air in the line. Refer to transducer manufacturer’s instructions on
how to remove trapped air from the transducer.
4.
Connect the patient catheter to the pressure line.
5.
Open the dome stopcock to room air.
6.
Zero the transducer. See
“Starting with accurate values” on page 14-4
.
7.
Open the dome stopcock to pressure catheter and check the quality of the waveform.
WARNING
Mechanical shock to invasive blood pressure transducer may cause severe
shifts in zero balance and calibration, and cause erroneous readings.
2
3
4
5
1
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......