B30 Patient Monitor
4-12
Recording time is 30 seconds consisting of 12 seconds recording from the recorder memory
and 18 seconds real-time recording. The alarm source is always marked to the alarm
recordings. The following alarms start the recording:
Showing alarm history
•
Press the
Pt. Data & Trends
key and select
Alarm History
. This displays a list of the
last 20 alarms that have reached the yellow or the red alarm level. The time and type of
occurrence are displayed next to the alarm list items.
Alarm
Recorded parameters
Asystole
Tachy/Brady
Art High/Low
V Fib
V Tachy
V Run >3
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
ECG1 + Art waveforms, 25 mm/s
Alarming level
Alarms which start the recording
Severe
Asystole, Tachy/Brady, Art High/Low, V Fib, V Tachy
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......