System description
2-33
Temperature
1
Measurement range:
10 to 45 °C (50 to 113 °F)
Measurement accuracy:
±0.1 °C (25 to 45 °C)
Probe type:
Use only GE Healthcare temperature probes or defibrillator-proof
YSI 400 series probes.
Temperature self-check:
At start-up and then every 10 minutes.
Time constant of temperature probes:
Reusable skin temperature probe:
3 s
Reusable adult central temperature probe:
6 s
Reusable pediatric central temperature probe:
4 s
Disposable skin temperature probe:
3 to 6 s
Disposable central temperature probe, 12F:
5 to 8 s
Disposable central temperature probe, 9F:
5 to 8 s
NIBP
Measurement range:
adult 25 to 260 mmHg (3.3 to 34.7 kPa)
child 25 to 190 mmHg (3.3 to 25.3 kPa)
infant 15 to 140 mmHg (2.0 to 18.7 kPa)
Typical measuring time:
adult less than 30 s, infant less than 25 s
Overall system accuracy:
Meets or exceeds SP10-2002 AAMI standards
2
The ESU does not cause a burn hazard through the NIBP cuff, because there is no electrical connection
between the cuff and the NIBP measuring electronics.
NOTE: NIBP measurement is intended for patients who weigh 5 kg (11 lb) and up.
NOTE: The cuff pressure measurement range is equal to cuff nominal and cuff
indication ranges.
Pulse oximetry
SpO
2
1
Automatic scaling of plethysmographic waveform.
Measurement and display range:
0 to 100 %
Calibration range:
70 to 100 %
Calibrated against functional oxygen saturation.
Measurement accuracy
3
:
100 to 70 %, ±2 digits
69 to 0 %, unspecified
Display resolution:
1 digit (1% of SpO
2
)
Wavelength of SpO
2
probe LEDs:
Infrared LED
940 nm
Red LED
660 nm
Maximum energy of SpO
2
probe LEDs:
Infrared LED
42 µJ/pulse
Red LED
62 µJ/pulse
Pulse rate:
Measurement and display range:
30 to 250 bpm
Measurement accuracy
4
:
±5 % or ±5 bpm, whichever is greater
1 The isolation barrier capacitance in the module has been minimized to reduce the hazard of burns in the event of a
defect in the ESU return electrode connection.
2 According to SP10-2002 AAMI 4.4.5.2.B, Intra-arterial method as the reference standard, mean difference of the test
system and the comparison system shall be ± 5 mmHg or less with standard deviation of 8 mmHg or less
3 Accuracy is based on deep hypoxia studies with volunteered subjects during motion and non-motion conditions over a
wide range of arterial blood oxygen saturations as compared to arterial blood CO-Oximetry. Accuracy may depend on
the sensor used, please refer to the instructions for use in the accessory package. The accuracy is expressed as rms. This
means that approximately two-thirds of the data will fall within the accuracy range.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......