![background image](http://html1.mh-extra.com/html/ge/b30/b30_users-reference-manual_4014252232.webp)
B30 Patient Monitor
17-6
Calibrating
NOTE: Ensure that the calibration gas and regulator are functioning properly before
calibration. Perform annual maintenance of the regulator as required.
NOTE: Do not wash or disinfect calibration gas sampling lines.
The airway module should be calibrated once every six months or whenever there are
indications of errors in the gas readings.
Calibrate the gas measurement with the GE Healthcare calibration gas. Do not use any other
calibration gases.
−
Use the regulator 755534 or equivalent.
−
Use the gas 755580 only and set the CO
2
concentration to 20%.
If you do not use the recommended calibration gases, the calibration does not succeed.
During gas calibration, % units are always used for CO
2
regardless of selected measuring units.
Figure 17-5 Attaching regulator to calibration can
1.
Attach the regulator to the gas container.
2.
Attach a new sampling line to the water trap. Connect the loose end of the sampling line
to the regulator on the gas container.
Figure 17-6 Connecting sampling line to the gas valve and feeding gas
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......