Troubleshooting
10-5
RAM error
Faulty RAM circuit on the CPU board.
•
Contact authorized service
personnel.
Recorder module removed
There is no recorder module.
•
Reconnect the N-FREC or N-FCREC
module if you need a recorder.
Replace Battery A
Replace Battery B
There is hardly any charge left in one
of the batteries. Also check the symbol
on screen.
•
Replace the battery as soon as
possible.
Replace D-fend
Water trap is partially blocked. This
decreases air flow.
•
Replace the water trap.
Sample line blocked
The sampling line inside or outside the
monitor is blocked, or the water trap is
occluded.
•
Change the sampling line and
water trap.
Select inflation limits
You are using a hose without an
automatic identification. When you try
to start the measurement, the monitor
goes automatically to the selection
NIBP Setup
-
Inflation Limits
.
•
Select appropriate inflation limits.
NOTE: Auto option is not available
for these hoses.
Sensor INOP
The gas measuring sensor is
inoperative or the temperature in the
module has increased.
•
Contact authorized service
personnel.
Small resp curve
Signal amplitude is less than 0.4 ohms.
SpO2 probe off
The finger or ear lobe may be too thin
or sensor halves are not aligned.
•
Check connection between sensor
and patient.
•
Replace the sensor.
SRAM error
Faulty SRAM circuit on the CPU board.
•
Contact authorized service
personnel.
Tachy
HR is equal to or above the higher
alarm limit.
•
Check the patient status.
Temperature error
The self-check has found an error in
the temperature measurement.
•
Contact authorized service
personnel.
Unable to measure Dia
Accurate diastolic pressure not
achieved because of artifacts, weak
pulsation etc.
•
Check the patient status.
•
Check the NIBP cuff placement.
•
Perform a new NIBP measurement.
Unable to measure Sys
Systolic pressure probably higher than
maximum inflation pressure, or
artifacts interfere in the systolic area.
•
Check the patient status.
•
Check the inflation limits.
•
Perform a new NIBP measurement.
Unstable zero pressure
Pressure is unstable at start of the
measurement.
•
Calm the patient and retry.
Message
Explanation
What to do
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......