Airway gas (CO2)
17-3
Patient connections
1.
Insert the N-FC or N-FCREC in the monitor. Align the module with the insertion guides and
push until it clicks.
2.
Check visually that the airway adapter connections are tight and that the adapter is
operating properly, then attach it to the patient.
3.
Make sure that the water trap container is empty and properly attached. The water trap
should be changed between patients and emptied whenever it is more than half full.
WARNING
Handle the water trap and its contents as you would any body fluid.
Infectious hazard may be present.
4.
Connect the gas sampling line to the sampling line connector on the water trap.
5.
Turn on the monitor. The monitor performs a self-check.
6.
Before connecting the patient, wait until the message ‘Calibrating gas sensor’ disappears.
Then make the patient connections as described below and connect the sampling line to
the airway adapter.
7.
Position the adapter with sampling port upwards. This prevents any condensed water
from entering the sampling line.
If N-FCREC or N-FC is used with O
2
and/or N
2
O contents higher than 40%, make sure that
FiO2
Level
and
N2O Level
are set accordingly through
Airway Gas
-
CO2 Setup
to enable O
2
and/
or N
2
O compensation.
Take the gas sample as close to the patient’s airway as possible, as shown in the illustrations,
and connect the sampling line to the patient’s airway adapter.
Position the adapter’s sampling port upwards to prevent any condensed water from entering
the sampling line.
NOTE: The message ‘Sampling line blocked’ may result if you attach the sampling line
to the water trap after turning on the monitor.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......