Pulse oximetry
12-5
Pulse oximetry menu
Pleth Scale
If the pleth scale selection is AUTO, the scale of the plethysmographic waveform display is
automatically set during ‘Pulse Search’. The scale is then kept constant throughout the case to
enable easy detection of changes in the patient's perfusion. To adjust the pleth scale further,
select one of the following:
2
,
5
,
10
,
20
,
50
,
Auto
.
With automatic
AUTO
scaling mode, the scale changes automatically if the amplitude of the
pleth waveform either exceeds the current scale or falls below the maximum value of the next
lower scale by 10% for 30 seconds or more. When the scale changes, the message ‘Scale
changed’ is displayed.
The scale indicator number is displayed on the left side of the waveform.
SpO
2
Response
Selects the SpO
2
averaging time. The selections are
Normal
: 12 seconds (default setting),
Fast
:
3 seconds.
Beat Sound Volume
Adjusting the SpO
2
beat volume also affects the beat volume of ECG.
When SpO
2
is monitored, the monitor provides a variable pulse beep so that the tone of the
pulse beep rises with increasing oxygen saturation and falls as saturation decreases. You can
select the volume from 1 (soft) to 10 (loud). With selection 0 there is no audible sound.
HR Source
Selects the heart rate source. If the ECG signal is affected by too much noise for a reliable heart
rate calculation, heart rate can also be calculated from invasive pressure (Art) or
plethysmographic pulse waveform
Pleth
. The selected heart rate source is displayed above the
numerical display of the heart rate. The color of the heart rate is the same as that of the source
parameter.
The
AUTO
selection priorities for heart rate calculation are: ECG (the lead with highest R-wave),
pressure (Art) and plethysmographic pulse waveform.
Others
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......