B30 Patient Monitor
16-4
Connecting the cuff hose
1.
Connect the cuff hose to the NIBP cuff by placing the opposite connectors in contact and
locking them together.
2.
Plug the NIBP cuff hose to the module.
NIBP Setup menu
Ready Prompt
Ready prompt gives an audible tone when the NIBP measurement is ready. Adjust the volume
of the beep tone from 1 (soft) to 10 (loud), or to 0 (OFF.)
Inflation Limits
When this selection is ‘Auto’, the monitor automatically identifies the cuff hose and selects the
right inflation pressure and alarm limits for adults. For children, and when using infant hoses
without identification, set the limit manually to
Child
or to
Infant
.
This selection allows you to override the automatic safety limit feature for the hose/cuff being
used.
The selections are:
Auto, Infant
,
Child
and
Adult
. The selection AUTO is not available when
using hoses without identification.
NOTE: You cannot select adult limits with an infant cuff hose.
NOTE: When using very large adult cuffs, use ‘Adult’ limits to prevent ‘Cuff loose’
message from displaying.
Calibration Check
Enables the calibration check with an external manometer:
Remove cuff from the module connector before entering the menu. When the menu is
displayed, attach an external manometer with pump to the connector. Pump approximately
200 mmHg and compare the readings of the manometer and screen. If the difference is more
than 4 mmHg, calibration by authorized service personnel may be necessary.
Previous Menu
Returns to the
NIBP
menu level.
NIBP
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......