B30 Patient Monitor
10-4
Lead changed
The monitor automatically switches
the ECG1 waveform selection to a
measurable ECG Lead (I, II, III, aVR, aVL,
aVF or V) if the current ECG1 is not
measurable. Note that the ECG
waveform changes according to the
lead it is measured from.
•
Check the lead.
Leads off
ECG trunk cable or any of the leads is
disconnected. Offset voltage between
two electrodes is too high.
•
ECG: Reconnect the disconnected
trunk cable, electrode or leadwire.
•
ECG: Change the trunk cable,
leadset and module.
Network down: xxx
Network cable is not connected.
The Central is shut down.
•
Check the network cable.
•
Check the Central.
NIBP manual
Autocycling mode is interrupted
because of an air leak or loose cuff.
•
Check the NIBP setup and restart
autocycling.
No battery backup
•
Replace the batteries.
Noise
Unreliable HR calculation or distorted
waveform, possibly during
defibrillation or because of motion
artifacts.
•
ECG: Check the patient status.
•
ECG: Check the electrodes.
•
Ensure that the patient is not
shivering.
•
Poor electrode quality or wrong
positioning.
•
Change the lead.
•
Remove the ECG cable from the
connector and reinsert it.
No P1 transducer
Transducer or channel x cable
disconnected.
•
Connect the transducer or the
cable.
No SpO2 probe
There is no SpO
2
probe.
•
Check connection between the
SpO
2
sensor and module.
No SpO2 pulse
No SpO
2
pulse can be detected.
•
Try other measuring sites.
Poor signal
Signal is not sufficient.
•
SpO
2
: Change the measuring site.
Printer error
Printer is not working properly.
•
Check that the network printer is
operational.
Printing...
Printing network printer has started.
•
Please wait until the printing is
finished.
Printing ready
Monitor has completed sending
printing data to the printer.
•
Please wait until the printing is
finished.
PVC rate limit changed
Limit has changed on the Arrhythmia
Workstation
•
Check the limit.
Message
Explanation
What to do
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......