![background image](http://html1.mh-extra.com/html/ge/b30/b30_users-reference-manual_4014252198.webp)
B30 Patient Monitor
14-6
Determining pressure values visually
By moving the horizontal cursor across the pressure waveform, you can get accurate pressure
values at selected points. This may be useful, for example, if the patient's breathing pattern is
irregular. The cursor is not available for pressures shown with a combined scale.
1.
Press the
Invasive Pressures
key.
2.
Select
Px Setup - Px Cursor
.
3.
Move the cursor up or down by turning the ComWheel. Every time the cursor is moved,
the time (hours and minutes) and pressure values are displayed on the screen. This way,
you can keep track of the changes made.
4.
You can remove the cursor by selecting
Remove cursor
. Note that if the cursor is not
removed, it remains visible in the Normal Screen.
Labeling channels
The label of the pressure channel sets its display scale, color, filter, alarm source and alarm
limits. The label descriptions are preconfigured.
To change the label:
1.
Press the
Invasive Pressures
key.
2.
Select
P1 Setup
.
3.
Select
Label
.
The labels are the following:
P1, P2
Standard labels
Art
Arterial pressure
CVP
Central venous pressure
PA
Pulmonary arterial pressure
RAP
Right atrial pressure
RVP
Right ventricular pressure
LAP
Left atrial pressure
ICP
First intracranial pressure
ABP
Arterial blood pressure
Both Art and ABP labels are available for situations when two arterial lines are desired but you
want to use different settings or alarm labels.
Invasive
Pressures
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......