![background image](http://html1.mh-extra.com/html/ge/b30/b30_users-reference-manual_4014252215.webp)
Non-invasive blood pressure
16-1
16 Non-invasive blood pressure
Overview
The non-invasive blood pressure (NIBP) measurement uses the oscillometric measuring
principle. The cuff is inflated with a pressure slightly higher than the presumed systolic
pressure, then slowly deflated at a speed based on the patient’s heart rate, collecting data
from the oscillations produced by the pulsating artery. Based on this data, the unit calculates
values for systolic, mean and diastolic pressures.
Blood pressure measurements determined with this device are equivalent to those obtained by
an intra-arterial blood pressure measurement device. Multiple intra-arterial sites were used.
You can set the NIBP module into an automatic cycling mode to make measurements at
desired time intervals. You can also measure NIBP continuously for five minutes in STAT mode
or take separate single measurements.
NOTE: Intervals below 10 minutes and STAT measurements are not recommended for
extended periods of time.
WARNING
The NIBP measurement is indicated for patients weighing over 5 kg (11 lb).
Module description
NOTE: In the monitoring system, use only one module at a time for measuring NIBP.
Figure 16-1
Modules measuring non-invasive blood pressure
E-PSMW
E-PSMPW
NIBP
connector
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......