B30 Patient Monitor
2-30
Performance
WARNING
Operation of the monitor outside the specified values may cause inaccurate
results.
Any fluctuations within the specified limits do not affect the performance.
NOTE: Information in this section can be especially useful to clinicians.
Power supply
Rated voltages and frequencies:
100 to 240 V 50/60 Hz
Allowed voltage fluctuations:
±10 %
Maximum power consumption:
150 VA
Battery operation
Batteries:
Exchangeable lithium-ion, 2 pcs max.
Charging time:
2 hours per battery pack
Operation time:
up to 4.5 hours
Environmental conditions
Operating temperature:
normal operation:
+5 to +40°C (41 to 104°F)
while charging batteries:
+5 to +35°C (41 to 95°F)
Storage and transport temperature:
-20 to +60°C (-4 to 140°F)
Relative humidity:
10 to 90 % noncondensing
Atmospheric pressure:
670 to 1060 mbar (500 to 800 mmHg)
Alarm behavior
The maximum alarm delay of the alarm at the monitor signal output to network: 5 seconds
If the alarm mode is latched, the technical alarms are latched as well. This does not comply with the NIBP
(IEC 60601-2-30) and invasive pressure (IEC 60601-2-34) standard requirements.
Silencing alarms for 5 minutes does not comply with the SpO
2
(ISO 9919) standard requirements.
Defibrillator & IABP synchronization connector
Analog output
ECG:
From first user lead (ECG1)
Gain: 1
V/mV
±10%
Delay:
< 15 ms
DC offset:
±100 mV max.
Frequency response:
0.05 Hz to 40 Hz
Invasive blood pressure:
From pressure labeled ‘Art’
Gain:
10 mV/mmHg ±2%
Delay:
< 35 ms
DC offset:
±20 mV max.
Frequency response:
DC to 30 Hz
The pacemaker pulses have been replaced with 2 ms ±20% fixed digital pulses at the ECG analog output.
A device that fulfils the requirements of the IEC 60601-1 standard can be connected to the defibrillator &
IABP synchronization connector. There are no other limitations, because the signals of the connector are
galvanically isolated from patient applied part of the ECG and invasive blood pressure measurements.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......