Non-invasive blood pressure
16-9
Automatic NIBP double check
If the NIBP value exceeds the alarm limits, a new measurement takes place automatically
(immediately, when
Manual
measurement is selected, and after 30 seconds when
Auto
measurement is selected.) If the alarm situation persists, an alarm is given.
Functioning of NIBP measurement during arrhythmias
The functioning of the NIBP measurement has been tested by using Biotek BP Pump W (Non-
invasive blood pressure monitor tester with wrist cuff) in the presence of the following
arrhythmias (available from the Biotek BP Pump W): Tachycardia, Bradycardia, Premature
Atrial Contraction #1, Premature Atrial Contraction #2, Premature Ventricular Contraction and
Atrial Fibrillation and PVCs.
The test results indicate that the presence of some arrhythmias during measurement may
increase the time required for the measurement.
Checklist
Check that:
•
The hose is correct: black NIBP hose for adults and children and light blue for infants.
•
The O-ring on the hose connector is intact.
•
The connector is firmly pushed inside the cuff tube.
•
The NIBP hose is properly connected to the module and will not detach if pulled.
•
The NIBP cuff is correct for the patient size.
•
There are no holes or cracks in the cuff bladder or cuff tube.
•
The symbol indicating the center of the bladder is over the artery.
•
All residual air is squeezed out of the cuff before wrapping it around the arm.
•
The cuff is not loose.
•
The cuff is at heart level.
•
The cuff tubes or NIBP tube are not kinked or squeezed together.
•
If leak is suspected, start venous stasis and check that the pressure is stable during stasis.
•
Non-invasive blood pressure
is selected to be displayed through
Monitor Setup
-
Screen Setup
.
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......