B30 Patient Monitor
14-4
Starting with accurate values
Pressure transducers generally produce a small signal even when no pressure is applied to
them. It is necessary to zero the monitor with the transducer to establish an accurate electrical
zero point.
Also, the position of the transducer effects the accuracy of the measurement. An error of 10
mmHg of static pressure is introduced for every 13.6 cm (5.4 inches) difference in height
between the mid-heart and the transducer.
To zero the transducer, open the transducer to air and do one of the following:
•
Press the
Zero Px
module key or the
Zero ALL
side panel key. This starts zeroing
immediately.
NOTE: Selecting
Zero ALL
does not zero ICP. Zero it separately.
•
Press the
Invasive Pressures
key
on the Command Board and select
Zero Pressures
.
During the zeroing process, the message ‘Zeroing’ is displayed. After the transducer is zeroed,
the message ‘Zeroed’ is displayed in the digit field. After each channel is zeroed, the time of
zeroing is displayed in the menu.
NOTE: Check zero level after power interruptions.
NOTE: Invasive pressures need to be zeroed after reconnecting the pressure
transducer or cable, and whenever the patient’s position is changed. If all channels
have not been zeroed, the message ‘InvBP not zeroed’ appears. However, the alarms
advance to yellow and red levels regardless of zeroing.
Invasive
Pressures
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......