-74-
(5) Attach the electrode cover.
(6) Close the wind protector.
(7) Turn the splicer on.
Objective lens
Electrode
Front
Rear
A
A
Electrode Flange
Electrode
4. Resetting the Arc Count
Reset the arc count to zero. For details of this operation, refer to the Arc
Count Clear Function of the maintenance menu, section 5.2.
5. Examination
(1) Turn the power switch on, and make sure that there is no dust or dirt
on the monitor screen.
(2)
Press the "ARC" key several times to check the arc discharge.