-32-
4. Press the "SET" key
The splicer operates cleaning ARC, initial
gap setting and cleave angle check. "CHECK
FIBER" message is displayed if "PAUSE1" is
"ON" in the "OPTION SET." menu.
(Display 2)
[NOTE] The Cladding axial offset value is
display if "DATA" is "ON" of the
"OPTION SET." menu. Refer to Section
5.2.
5. Automatic Splicing Procedure
(1) Press the "SET" key to start splice process
again if the splicer in PAUSE state
("PAUSE1").
(2) The splicer aligns the clad in both X and Y
field, and displays "XY ALIGNING"
massage.
(3) The splicer aligns
θ
axis, and displays
"PM ALIGNING" message. The splicer
beeps 3 times after
θ
axis alignment is
finished, and will be in a PAUSE state if
“PAUSE2” is ON.
(4) Press the “SET” key if the splicer is in a
PAUSE state. The splicer aligns the cores
by RIDS. (Display 3)
(5) The splicer will be in a PAUSE state after
all alignments are finished. (Display 4)
[NOTE] In (3) and (5) process, the splicer will
be in a PAUSE state if "PAUSE2" is
"ON" of the "OPTION SET" menu.
The splicer displays "PAUSE"
message on the monitor during
PAUSE state. (Display 4) Refer to
section 5.2.
P2-PP
225
0.7 0.2
CHECK FIBER (X)
Display 2
P2-PP
XY ALIGNING (Y )
Display 3
P2-PP
PAUSE (Y)
Display 4
Summary of Contents for FSM-20PMII
Page 1: ...II ...
Page 2: ... 1 1 GENERAL 1 GENERAL 1 GENERAL 1 GENERAL ...
Page 4: ... 3 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION ...
Page 15: ... 14 4 OPERATION 4 OPERATION 4 OPERATION 4 OPERATION ...
Page 18: ... 17 Hear is Flowchart page 1 ...
Page 19: ... 18 Hear is Flowchart page 2 ...
Page 42: ... 41 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS ...
Page 58: ... 57 7 OHTERS 7 OHTERS 7 OHTERS 7 OHTERS ...
Page 66: ... 65 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE ...
Page 80: ... 79 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING ...
Page 88: ... 5 6 2 0 0 0 2 0 0 0 5 5 3 1 2 1 26 6 20 5 2 0 0 2 0 0 9 A 2 6 3 0 0 0 0 2 3 A 2 6 8 0 0 1 ...