-82-
Message
Reason
Remedy
ERROR 11
UNDER RANGE
Detected level of power
meter is too small.
In the case of P2 or P3 mode, the
power of the incident light is under the
detectable level of the power meter.
Check fibers and connection. Set fiber
again if the offset of fiber axis is large.
ERROR 12
GPIB INACTIBE
Power meter is not
active.
In the case of P2 or P3 mode, the
power meter is not active or GP-IB
cord is not connected properly in the
automatic aligning sequence. Check
the power meter is on and GP-IB
connection. Check the address switch
of the power meter is TALK ONLY.
ERROR 13
LEVEL
UNCHANGED
The level of power meter
doesn’t change.
X,Y, PL or PR motor doesn’t move
normally.
Consult the sales agency or factory.
ERROR 14
COVER OPEN
Wind protector is open
during PL or PR motor
moving.
Close the wind protector and press
RESET key.
Summary of Contents for FSM-20PMII
Page 1: ...II ...
Page 2: ... 1 1 GENERAL 1 GENERAL 1 GENERAL 1 GENERAL ...
Page 4: ... 3 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION 2 CONSTRUCTION ...
Page 15: ... 14 4 OPERATION 4 OPERATION 4 OPERATION 4 OPERATION ...
Page 18: ... 17 Hear is Flowchart page 1 ...
Page 19: ... 18 Hear is Flowchart page 2 ...
Page 42: ... 41 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS 5 MENU COMMANDS ...
Page 58: ... 57 7 OHTERS 7 OHTERS 7 OHTERS 7 OHTERS ...
Page 66: ... 65 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE 8 MAINTENANCE ...
Page 80: ... 79 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING 9 TROUBLE SHOOTING ...
Page 88: ... 5 6 2 0 0 0 2 0 0 0 5 5 3 1 2 1 26 6 20 5 2 0 0 2 0 0 9 A 2 6 3 0 0 0 0 2 3 A 2 6 8 0 0 1 ...