89
Splice Mode
Parameter
Description
Left Fiber / Right Fiber [FSM-100P/P+ Only]
Correction
Angle
The splicer rotates both left and/or right fibers through an additional
correction angle if one is programmed for either fiber.
NOTE:
A correction angle may sometimes be useful to improve
alignment accuracy for a particular type of PM fiber or for a particular
fiber production lot.
IPA Settings [FSM-100P /P+ Only]
Rotation Step
Angle
Sets rotation step angle of IPA operation.
Can be selected the resolution between 1 and 10 degrees.
Total Rotation
Angle
Sets total rotating angle of IPA operation.
Can be set to [180°] or [360°].
Left Fiber / Right Fiber [FSM-100P/P+ Only]
Method
Can be set to [Auto], [Diff.Fiber] or [Same Fiber].
Reference
Model
Can be set to [Find Best],
[*PANDA],
[*BOWTIE],
[*PANDA-S1],
[*PANDA-S2], or [IPA Data] registered by [Fiber Data Learning].
For the details, see section [
].
Additional θ Alignment [FSM-100P/P+ Only]
θ Motor Drive
[EV]
If [θ Motor Drive [EV]] is set to "ON", "θ Motor Drive [EV]" screen
appears after gap set or θ alignment. At "θ Motor Drive [EV]" screen
rotational position of the fiber can be adjusted manualy while the fiber
end is shown on the screen.
Focus Control
If [Focus Control] is set to "ON", focus is adjusted automaticaly when
the "θ Motor Drive [EV]" screen appears or the view for the fiber is
changed on "θ Motor Drive [EV]" screen.
•
The IPA Settings of [θ Alignment] and [Angle Offset Estimation] are common
settings. If the IPA Settings of [θ Alignment] are changed, the settings for
[Angle Offset Estimation] are also automatically changed to the same settings.
•θAngle Adj. can be used, even if [θ Align] is “OFF”.