105
Splice Mode
Parameter
Description
Error Limit Threshold
Cleave Angle
Limit
An error message is displayed if the cleave angle of either the left or
right fiber ends exceeds the selected threshold (cleave limit).
“OFF”: Does not make judgment on cleave angle.
Cleave Shape
Error
Set the threshold of Cleave Shape Error limit.
“OFF”: Does not make judgment on fiber angle.
Fiber Angle
Limit
An error message is displayed if the bend angle of the two fibers
spliced exceeds the selected threshold (Fiber Angle Limit).
“OFF”: Does not make judgment on fiber angle.
Loss Limit
An error message is displayed if the estimated splice loss exceeds the
selected threshold (loss limit).
“OFF”: Does not make judgment on splice loss.
Angle Offset
Limit
If [Angle Offset Method] is ON.
Set the threshold of estimated Angle Offset error. When the estimated
Angle Offset is higher than threshold, an error message is displayed.
“OFF”: Will not scrutinize estimated Angle Offset.
Arc Center
Limit
An error message is displayed if the center of the arc discharge exceeds
the selected threshold (Arc Center Limit).
If the value increases, it is easier to detect.
Bubble
Sensitivity
Set the bubble error threshold. When the bubble is higher than the
threshold, an error message is displayed.
Fat Sensitivity
Set the fat error threshold. When a bulge in the splice is higher than the
threshold, an error message is displayed.
Thin Sensitivity
Set the thin error threshold. When a thin area of the splice is higher
than the threshold, an error message is displayed.
Hot Spot
Sensitivity
An error message is displayed if the arc discharge is abnormal (Hot
Spot Sensitivity).