7
Quick Start Guide
1-2.
FSM-100P / 100P+
The splice setting of the FSM-40F/PM series and the FSM-45F/PM series can be
converted into the FSM-100 series. Please refer to the Excel file that exists in CD.
Fiber Combination
PANDA
Both sides
Excluding
PANDA
Both and Either
side
LDF
PMF
PM Fiber
Clad Diameter
125µm
125µm
400µm
125µm
Recommended
Splice Mode
PANDA 1
PM AUTO 1
PANDA 4
PANDA8
Auto Stuff Control
ON
ON
ON
ON
Electrode Gap
1mm
1mm
3mm
1mm
V Height Shift
0µm
0µm
0µm
0µm
Arc Calibration
Method
See Section 2-1
STD
(Standard)
STD
(Standard)
SP
(Special)
Depend on
fiber type
Alignment Mode
Clad
Clad
Clad
Clad
Focus Settings
0.32
Auto
Auto
Auto
Fiber Type Compare
See Section 3-1
-
Auto Arc Calibration
See section 2-2
OFF
OFF
OFF
OFF
Auto Time Comp
See Section 2-6
OFF
OFF
OFF
OFF
Arc Center
Compensation
See Section 4
OFF
OFF
OFF
OFF
θ
Alignment
See Section 2-4
PAS
IPA
PANDA
EV
Axis Offset
Clad
Clad
Clad
Clad
Core Bending
OFF
OFF
OFF
OFF
MFD Mismatch
OFF
OFF
OFF
OFF
Note
The PANDA
fiber is spliced
accurately at
the shortest
time.
If clamping on
the coating,
alignment time
is longer.
All PM fibers can
be aligned by
using the data of
the fiber prepared
beforehand,
although it is
longer.
Moreover, it is
applied to splice
dissimilar fiber
types.
This mode is
for Large
Diameter
Fiber splicing.
Please
execute SP arc
calibration.
This mode is
for PM Fiber
splicing with
end-view
observation
system.
Note