68
Arc Calibration
Item
Arc Calibration
Special Arc Calibration
Arc Calibration
Method
STD (Standard)
If STD is not selected, it doesn’t
operate.
SP (Special)
If SP is not selected, it doesn’t
operate.
Electrode Gap
1.0 mm
Any Edited Value
V Height Shift
0 µm
Any Edited Value
Clad Diameter
125 µm
Any Edited Value
Arc Center
Compensation
Calibration Provides Correction
Calibration Provides Correction
Calibration Result
Arc power is calibrated. The result is
applied to
all
splice modes that utilize
the standard [STD] Arc Calibration.
Special Arc Calibration [SP] result is
used to calculate and set appropriate
arc power
only for the selected splice
mode.
Further arc power optimization
may be required for best performance.
[Electrode Gap]
[V Height Shift]
[Clad Diameter]
[Electrode Gap]: 1 mm only
[V Height Shift]: 0 µm only
[Clad Diameter]: 125 µm only
Each parameter is limited.
[Electrode Gap]: selectable
[V Height Shift]: selectable
[Clad Diameter]: selectable
Each parameter is programmable