81
Splice Mode
In other splice modes in the user-selectable database, the user can select one from a
series of factory-set splicing modes for various splicing combinations. Below are the
descriptions of the various parameters used in these modes.
Parameter
Description
Fundamental Settings
Fiber Type
A list of splice modes stored in the splicer database is displayed.
Upon inputting the appropriate mode, the selected splice mode
stored in database area is copied to a selected splice mode in
user-programmable area.
※
The parameter doesn't exist according to the model because this
chapter explains in the Full mode.
Mode Title 1
Title for a splice mode expressed in up to nine characters.
Mode Title 2
Detail explanation for a splice mode expressed up to 15 characters.
Operating Mode
[Full] [Basic] [Fiber Shaping] selected [Edit Splice Mode] of The
composition of the parametrical expression changes.
Use [Full] mode when you want to do a detailed
setting
The Basic mode doesn't include the Special
Function mode.
The Fiber Shaping mode can be used for the
roller end face processing etc.
Auto Stuff Control
[Prefuse Power] and [Prefuse Time] change to the recommended
value when [ON] is selected and the splicer controls [Stuff Speed]
at the optimum speed.
This function doesn't operate when [OFF] is selected.
White display indicates [ON]. Yellow display indicates [OFF].
• The menu composition changes when the mode is
changed and selected with Operating Mode.
Edit
display
•
The splice loss will be inconsistent for specific types of fiber