88
Splice Mode
Parameter
Description
Left Fiber / Right Fiber [FSM-100P/P+ Only]
Alignment
Mode
When [PAS], [Powermeter], [EV] or [EV Lock] is selected as the
[Alignment Method], the following screen image is displayed to enable
selection of an [Alignment Mode] to correspond to the [Alignment
Method].
Select [Panda], [IPA], [B1500T], [Angle], [D-Shape], [Panda2], [Ecc]
or [Manual] when [PAS] is selected as the [Alignment Method].
Select [Max], [Min] or [Manual] when [Powermeter] is selected as the
[Alignment Method].
Select [Auto], [Panda], [Bowtie], [Tiger], [PCF], [MCF], [PM-PCF],
[DC-Panda], [Octagon] or [Manual] when [EV] is selected as the
[Alignment Method].
Select [Auto], [Panda], [Bowtie], [Tiger], [PCF], [MCF], [PM-PCF],
[DC-Panda], or [Octagon] when [EV-Lock] is selected as the
[Alignment Method].
NOTE 1:
[EV] and [EV Lock] are available only with the FSM-100P+
NOTE 2:
If [PAS] is selected as the [Alignment Method], many Panda
type fibers as listed below are very quickly and accurately aligned using
[Panda] as the [Alignment Mode]. Some other low-contrast Panda fibers
as well as fibers listed below under “IPA” are best aligned using [IPA] as
the [Alignment Mode].
Panda
Fujikura
SC37-PS-U40A
、
SC48-PS-U40A
SM63-PS-U40A
、
SM85-PS-U40A
、
SM98-PR-U25A
SM98-PS-U40A
、
SM13-PR-U25A
、
HA13-PS-U25A
SM13-PS-U40A
、
SM15-PR-U25A
SM15-PS-U40A
、
DS15-PS-U40A
、
ED15-PS-U25A
Nufern
PM980
、
PM14XX
、
PM1550
、
PM980C
、
PM14XXC
IPA
Fibercore
HB1500, HB1500T, HB980T
StockerYale PMF-1550-B1, PMF-820-B1 PZF-820-B1
The alignment
method is shown on
the display.
If the alignment
mode is OFF,
“OFF” is shown on
the display.