164
Maintenance Menu
9.
Motor Calibration
(
ZL/R, X/Y, FcsX
・
Y, θL/R, Sweep
)
Motors are adjusted at the factory before shipping. However, these settings could have
changed due to various reasons. This function automatically calibrates the speed of all
motors. (Z, XY, Focus, θ, Sweep)
Operation Procedure
(1)
Select the [Motor Calibration] in the [Maintenance Menu].
(2)
Load prepared fibers in the splicer and press ENT key.
(3)
Speeds for all motors are automatically calibrated. Upon completion,
[Maintenance Menu] is displayed.
• Perform this function when “Fat” or “Thin” error has occurred, or fiber
aligning or focusing is taking too much time.