73
Arc Calibration
(2)
Place prepared fibers in the splicer.
(3)
Press ENT and perform the following.
During additional electrical discharges, the electrical discharge power is
adjusted after measuring the round end face shape.
Cleave angle threshold does not link to the parameter “Cleave Limit” in
splicing modes. Cleave angle threshold is independently set for arc
calibration. See the [
] to change cleave angle
threshold.
•Use the assigned fiber.
•Use glass clamping.
•Use cleaned fibers for arc calibration. Dust affects arc calibration.
Right and left fibers
advance.
The electrical discharge
is observed.