Operating Manual BKS600A
28
9.4
Description of the system parameters
The parameter changing mode is activated by pressing the
PARA
↵
key for 3 seconds. By
pressing the
PARA
↵
key again, the system parameters are selected (ref. also to fig. 22).
Language
Use:
This parameter stores the display language.
Range:
English, French, Italian, German
Measuring system
Use:
This parameter indicates the measuring system to be used.
Range:
Metric, US standard
Default:
Metric
Note:
(not used from the BKS600A)
Lowpass display
Use:
The electronic unit provides a lowpass filter to prevent noise which is added
to the integrated display. This parameter stores the cut off frequency. The
lower the cut off frequency, the more sluggish the output signal will be. Due
to this filter, the value shown in the display will be much more stable in the
case of high fluctuations of the force value.
The lowpass display filter is independent to the other filters.
Range:
0.1
to
10.0
Default:
1.0
Increment:
0.1
Unit:
[Hz]
Identifier
Use:
This parameter stores the ident number of the device when linked to
PROFIBUS, CAN-Bus resp. DeviceNet.
Range:
0
to
255
Default:
0
Increment:
1
Unit:
[-]
Baud rate
Use:
This parameter stores the speed of the serial interface (RS232). The other
settings are fixed: 8 data bits, even parity, 1 stop bit („8 e 1“).
Range:
2400, 4800, 9600, 19200
Default:
9600
Unit:
[Baud]