Page 3.43
=============================================================
QUERY WAVEFORM QW
-------------------------------------------------------------
Purpose:
Queries the trace data (administration and/or sample data)
related to the waveform from the ScopeMeter.
When a waveform is queried that is still under processing,
the processing is finished first (no half traces returned).
Command Syntax:
QW <trace_no>[,V|S]
<trace_no> = <decimal number>
<trace_no> Trace Source: (only for Fluke 19x)
---------------------------------------------
10 Scope mode:
Normal trace INPUT A
Min/Max trace INPUT A
Scope Record:
Min/Max trace INPUT A
11 TrendPlot 1:
Min/Max/Average trace
20 Scope mode:
Normal trace INPUT B
Min/Max trace INPUT B
Scope Record:
Min/Max trace INPUT B
21 TrendPlot 2:
Min/Max/Average trace
30 Scope mode, Mathematics
(not available in all versions):
Min/Max trace A+B, A-B or AxB
<trace_no> Trace Source: (only for Fluke 19xC)
---------------------------------------------
10 Scope mode:
Normal trace INPUT A
Min/Max trace INPUT A
Scope Record:
Min/Max trace INPUT A
11 TrendPlot 1:
Min/Max/Average trace
12 Scope mode:
Min/Max trace INPUT A ENVELOPE
13 Scope mode:
Min/Max trace INPUT A REFERENCE
20 Scope mode:
Normal trace INPUT B
Min/Max trace INPUT B
Scope Record:
Min/Max trace INPUT B
Summary of Contents for ScopeMeter 190 Series
Page 40: ...Page 3 29 Response Syntax acknowledge cr See an example for this command under QUERY PRINT QP...
Page 42: ...to the ScopeMeter...
Page 64: ......
Page 68: ...Resp Count Resp Count 2 PRINT x_divisions X Divisions...
Page 70: ...PRINT x_step X Step...
Page 72: ...nb nb ASC MID Resp Resp Count 1 1 ELSE...
Page 74: ......
Page 76: ...PRINT SumCheck so far MOD 256 SumCheck1 MOD 256 CLOSE END...
Page 78: ...END IF...
Page 80: ...END IF NEXT j...
Page 83: ...Resp Count Resp Count Sample Bytes...